Refer to the "see also" section of this page for more information regarding the tones. If the machine only says "Service Required", it is usually remstar error codes a fatal error. This will happen with water damage that has caused a
Respironics Cpap Error Codes
short circuit. Below are the error codes for the REMstar xPAP machines. Please note: these codes are for older
Remstar Shuttle Error Codes
pre-2007 REMstar machines. For the newer PR System One machines, go to Philips-Respironics error codes. Contents 1 E-0 - E-20 2 E-21 - E-40 3 E-41 - E-60 4 E-61 - E-80
Respironics Cpap Service Required
5 E-80 - E-86 6 See Also E-0 - E-20 Error Code Description Probable Cause Corrective Action E-0 NO_ERROR Reserved N/A E-1 ERR_SOFTWARE Reserved Replace Main PCA E-2 ERR_PROGRAM_CRC_FAILURE Defective FLASH chip Replace Main PCA E-3 ERR_EXTERNAL_RAM_FAILURE Defective RAM chip Replace Main PCA E-4 ERR_FIQ_STACK_OVERFLOW Defective RAM chip Replace Main PCA E-5 ERR_FIQ_STACK_UNDERFLOW Defective RAM chip Replace Main PCA E-6 ERR_NEST_STACK_OVERFLOW Defective RAM chip Replace philips respironics service required message Main PCA E-7 ERR_NEST_STACK_UNDERFLOW Defective RAM chip Replace Main PCA E-8 ERR_IRQ_STACK_OVERFLOW Defective RAM chip Replace Main PCA E-9 ERR_IRQ_STACK_UNDERFLOW Defective RAM chip Replace Main PCA E-10 ERR_TMR_STACK_OVERFLOW Defective RAM chip Replace Main PCA Error Code Description Probable Cause Corrective Action E-11 ERR_TMR_STACK_UNDERFLOW Defective RAM chip Replace Main PCA E-12 ERR_SVC_STACK_OVERFLOW Defective RAM chip Replace Main PCA E-13 ERR_SVC_STACK_UNDERFLOW Defective RAM chip Replace Main PCA E-14 ERR_THREAD_STACK_OVERFLOW Defective RAM chip Replace Main PCA E-15 ERR_ARM_UNDEFINED_INSTRUCTION_EXCEPTION Defective FLASH chip Replace Main PCA E-16 ERR_ARM_SWI_EXCEPTION Defective FLASH chip Replace Main PCA E-17 ERR_ARM_PREFETCH_EXCEPTION Defective FLASH chip Replace Main PCA E-18 ERR_ARM_ABORT_EXCEPTION Defective RAM chip Replace Main PCA E-19 ERR_ARM_RESERVED_EXCEPTION Defective FLASH chipDefective microprocessor Replace Main PCA E-20 ERR_AIC_DEFAULT_HANDLER Defective FLASH chip Replace Main PCA E-21 - E-40 Error Code Description Probable Cause Corrective Action E-21 ERR_AIC_SPURIOUS_HANDLER Defective FLASH or RAM chip Replace Main PCA E-22 ERR_CAL_DATA_CRC_FAILURE Improper calibrationDefective FLASH chip RecalibrateReplace Main PCA E-23 ERR_INVALID_DRIFT_INDEX_AT_CAL Improper calibrationOccluded or restricted pressure tubeDefective flow sensor RecalibrateReplace pressure tubeReplace Main PCA E-24 ERR_UNRECOGNIZED_CAL_VERSION_BY_SW Improper calibrationDefective FLASH chip RecalibrateReplace Main PCA E-25 ERR_DRIFT_FLOW_TOO_HIGH Occluded or restricted pressure tubeDefective flow sensor Replace Flow TubeReplace Main PCA E-26 ERR_UI_DISPLAY_FAILURE
to CPAP and Sleep Apnea (CLICK HERE TO READ POSTS) Post a reply Print View 32 posts • Page 1 of 3 • 1, 2, 3 qttrcoupe Posts: 1Joined: Wed Feb 21, 2007 10:12 amGender: Reply with replace main pca quote Remstar Error Codes by qttrcoupe on Wed Feb 21, 2007 10:20 am Does anyone philips remstar plus c-flex service required have a list of the error codes.. I have seen a couple on here I keep showing up and E-20 on my respironics bipap autosv error codes Remstar Auto M C-Flex... Not sure what it is. the machine still functions normally but I am growing concerned as it did it twice last night just before bed. Thanks Kyle Top Wulfman Posts: 12345Joined: Thu http://www.apneaboard.com/wiki/index.php?title=REMstar_error_codes Jul 07, 2005 3:43 pmLocation: Nearest fishing spotGender: Reply with quote by Wulfman on Wed Feb 21, 2007 11:33 am Kyle, The service manual says that the E-20 error is: ERR_AIC_DEFAULT_HANDLER Defective FLASH chip Replace main PCA (that's the main motherboard) You're going to have to send it in to the factory. These machines have a 2-year warranty. Hopefully it's not past that. Best wishes, Den (4) REMstar Autos w/C-Flex & (6) REMstar http://www.cpaptalk.com/viewtopic/t17622/Remstar-Error-Codes.html Pro 2 CPAPs w/C-Flex - Pressure Setting = 13 cm. "Passover" Humidification - ResMed Ultra Mirage FF - Encore Pro w/Card Reader & MyEncore software - Chiroflow pillowUser since 05/14/05 Top RB Reply with quote Error Codes by RB on Sat Nov 03, 2007 8:34 am Once again does anyone know where to find a complete listing of the error codes for the Remstar Auto? Not just one at a time but the complete list. Thank you Top Wulfman Posts: 12345Joined: Thu Jul 07, 2005 3:43 pmLocation: Nearest fishing spotGender: Reply with quote Re: Error Codes by Wulfman on Sat Nov 03, 2007 1:26 pm RB wrote:Once again does anyone know where to find a complete listing of the error codes for the Remstar Auto? Not just one at a time but the complete list.Thank you Yeah.....they're in the "Service Manual" for the machines. The manuals are in .pdf format. But.....just for YOU.....I've converted that section to text format and am posting the complete list for the Auto and Pro 2. Be sure to copy them to your computer so I don't have to do this AGAIN! REMstar Auto & Pro 2 error codes: Error Code - Description - Probable Cause - Corrective Action E-0 NO_ERROR •Reserved •N/A E-1 ERR_SOFTWARE •Reserved •Replace Main PCA E-2 ERR_PROGRAM_CRC_FAILURE
by R. Wilson at 2:39 PM 0 comments Friday, September 01, 2006 My Remstar Pro w/ CFlex started giving me an Error code, E-24. So I got an http://cpapfix.blogspot.com/ Rx from my doctor to "Repair or replace" my CPAP.Well my insurance would not pay to repair, or replace it. The best I could do was they would pay to "rent to purchase", but I have a $500 deductible, & a new machine is about $600. So I’d be paying almost all of it!So I did a Google search for “Remstar Pro E-24 error code" And came to this error codes forum....http://www.cpaptalk.com/viewtopic.php?t=11773&start=0&postdays=0&postorder=asc&highlight=&sid=014fbf557f7ee365f1a9bdc249e84fb1Where a post states..."An E-24 error on a REMstar Plus w/C-Flex is:Problem: Flow offsetCorrective Action: Check for leaks or kinked internal tubing.Replace Main PCA.Looks like you need to take/send it in to the DME/place where you bought it. "So I took my machine apart to look inside. I did find a hose that should've been connected to a plastic nipple that had broken off.A few drops of Super glue & philips respironics system Walla!Now, I Still had this error code showing up.So I find on E-bay people selling “Clinicians manuals” for about $5 each.(http://search.ebay.com/cpap-clinician-manual_W0QQfromZR8QQfsooZ1QQfsopZ1QQfsprZ1QQssPageNameZWD8B)The listings state they are “Home Care Provider Set up Instruction Manual” And“This is the manual that allows you to enter the Therapy Set up Menu to Change and Set Pressures, Adjust C Flex, Alarms, Display Lights, Elevation Settings and all other features.”I e-mail the seller & ask how many pages the manual is and if you need any other equipment to access & change setting such as the prescribed pressure?The response was that no other equipment is necessary; the manual tells you how to access the menus to change any and all settings. And it is 8 pages.So I start searching the web to find the manuals or instructions on how to do this.It didn’t take to long before I found an actual Respironics TECHNICIANS SERVICE MANUAL!!!157 Pages!!This thing tells you how to take the machine apart & completely replace any component! Along with how to access the therapy set up menu and change any and all settings!!!!!On page 34 it even states “to prevent patients from tampering with the settings, the directions for entering the therapy menu should not be revealed to the patient”.It also list